PORTABLE AFTER CATION CONDUCTIVITY
Image 1
Image 2

Description

This portable system is available in two configurations: as a Trolley Mounted System or as a Hand Carry (can easily be carried in the hand). It consists of an After Cation Conductivity Meter for offline measurement of After Cation Conductivity values of various Steam-Watersamples (Feed -Water, Main-Steam, Condensate, etc.) available in the SWAS room. In general, the ACC values of these samples range from 0.06 to 1 micro siemens/cm. The instrument shall contain one Cation Column with resin, Sample Flow Indicator with needle valve, one conductivity Cell/Sensor, and one conductivity Value Display Unit (can be operated by a power supply 220 VAC 50 Hz and with battery power).

Features

• Sample Isolation Valves ¼” OD

• Cation Column with Resin in H+ Form

• Rotameter with Needle Valve

• SS316, 1/4” Flow Through Chamber for Conductivity Sensor

• Conductivity Analyzer with Sensor

• Pressure Regulator Sample Tubing – ¼” OD 5 Meter

• UPS Power Supply (For 4 Hrs of operation)

• Sample Drain Point

• Flexible Tubing 4-5 meters, 1/4” OD

Specifications:

MeasurementCation Conductivity / Specific Conductivity
Accuracy< ± 1%
Response Time (90% of Full Scale)< 5 sec
Repeatability± 0.1 %
Range0.01 - 1 µS/cm Note: Multiple Ranges Available
Cell ConstantK = 0.01
Resolution0.001
Temperature CompensationAutomatic
Sample Flow Rate100-250 mL/min (max)
Sample Temperature-5 to 50°C
Analog Output Configuration4-20mA per measurement
Temperature Sensor and AccuracyPt-1000 RTD
CommunicationModbus
Cation Resin900 mL
DisplayBacklit LCD
Ambient Temperature± 10 - 45 °C
Storage and Transport Temperature± 10 - 50 °C
Power Supply24VDC / 110/230 VAC, 50Hz ±5% (User defined)
Humidity10-90% Non-Condensing
Ingress ProtectionIP54