We are available during working hours on the following numbers:
This portable system is available in two configurations: as a Trolley Mounted System or as a Hand Carry (can easily be carried in the hand). It consists of an After Cation Conductivity Meter for offline measurement of After Cation Conductivity values of various Steam-Watersamples (Feed -Water, Main-Steam, Condensate, etc.) available in the SWAS room. In general, the ACC values of these samples range from 0.06 to 1 micro siemens/cm. The instrument shall contain one Cation Column with resin, Sample Flow Indicator with needle valve, one conductivity Cell/Sensor, and one conductivity Value Display Unit (can be operated by a power supply 220 VAC 50 Hz and with battery power).
• Sample Isolation Valves ¼” OD
• Cation Column with Resin in H+ Form
• Rotameter with Needle Valve
• SS316, 1/4” Flow Through Chamber for Conductivity Sensor
• Conductivity Analyzer with Sensor
• Pressure Regulator Sample Tubing – ¼” OD 5 Meter
• UPS Power Supply (For 4 Hrs of operation)
• Sample Drain Point
• Flexible Tubing 4-5 meters, 1/4” OD
| Measurement | Cation Conductivity / Specific Conductivity |
|---|---|
| Accuracy | < ± 1% |
| Response Time (90% of Full Scale) | < 5 sec |
| Repeatability | ± 0.1 % |
| Range | 0.01 - 1 µS/cm Note: Multiple Ranges Available |
| Cell Constant | K = 0.01 |
| Resolution | 0.001 |
| Temperature Compensation | Automatic |
| Sample Flow Rate | 100-250 mL/min (max) |
| Sample Temperature | -5 to 50°C |
| Analog Output Configuration | 4-20mA per measurement |
| Temperature Sensor and Accuracy | Pt-1000 RTD |
| Communication | Modbus |
| Cation Resin | 900 mL |
| Display | Backlit LCD |
| Ambient Temperature | ± 10 - 45 °C |
| Storage and Transport Temperature | ± 10 - 50 °C |
| Power Supply | 24VDC / 110/230 VAC, 50Hz ±5% (User defined) |
| Humidity | 10-90% Non-Condensing |
| Ingress Protection | IP54 |